Infant Incubator

Brief introduction:
Servo-controlled by micro-computer; sufficient operating window can reduce the disturbance to the baby during breast-feeding; touch switch panel with LED digital display, set-temp, real temp is shown separately; multiply yourself by checking the alarm sound and look; heating shows power level of 000-100%, a crib can be pulled out from a side door.
Standard configure: Hood, infant bed, temperature controller, water tank, air filter, IV Pole, stand.
Technical Specifications:
Power supply: AC220V 50Hz Power input: 600 VA
Control of air temperature ranges? 25 ~ 37
Changes in temperature: the temperature uniformity of 0.5 d: d 0.8
Temperature rise time: 45min d
Noise level: 55dB A Worrying d: audible and visible alarm indication, and may show the real temperature and cut off power supply from the heater.
Over-temp alarm: when the temperature control d38
Deviation alarm: ± 3?
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or placed in the wrong place;
Fan alarm: when the fan is blocked, d1000rpm speed or stop working;
Power failure alarm: when there is no electricity or power cable disconnected.
Alarm system: when there is damage in E2ROM temperature control.
Package: 72Kg Gross Weight
Size: 1085mm × 640mm × 1065m